IM 701240-02E
Observing Waveforms
4
Press the
Initialize
soft key.
Initialization is executed.
Press here to set the SL1400 back to
the condition that existed immediately
before the initialization.
Note
• When you initialize, the display of all channels in which modules are installed is turned ON in the stopped condition.
• The items that cannot be initialized using the Initialize soft key are as follows:
Settings such as the date/time, recorder mode, language, communication parameters, and SCSI ID.
• To initialize all settings except the date/time, carry out the procedure below. You cannot restore the settings. The stored setup data
is also initialized (see section 13.1 in the user’s manual).
[Turn ON the power switch while holding down the RESET key]
Performing Auto Setup
Section 4.5, "Performing Auto Setup" in the user's manual
Vertical axis (voltage range), horizontal axis (record time), trigger, and other settings are automatically configured according
to the input signal. This is useful when you wish to quickly display the waveform or when you do not know the setup condi-
tions because the characteristics of the input signal are unknown.
1
Press
MENU
.
2
Press the
Next 1/2
soft key.
3
Press the
INITIALIZE/CAL
soft key.
4
Press the
Auto Setup
soft key.
Auto setup is executed.
Press here to set the SL1400 back
to the condition that existed
immediately before auto setup.
Waveform of the signal applied
to CH1
MENU