17-14
IM DLM3054-01EN
Power Measurement (Power Measurement)
Power Measurement1, Power Measurement2
Power of up to two circuits can be measured simultaneously.
Turning Power Measurement ON or OFF (Mode)
Set whether to make power measurements.
• ON: Power measurement is enabled.
• OFF: Power measurement is disabled.
•
and power measurement of the power supply analysis feature cannot be executed
simultaneously. If any of the power measurement items, Power Measurement1 to Power Measurement2, is
set to ON, the power supply analysis is set to OFF. If power supply analysis is set to something other than
OFF, all power measurements are set to OFF.
• For input channels that are assigned to power measurement and whose mode is set to ON, the following
waveform parameters cannot be set: Max, Min, P-P, Rms, Mean, Sdev, Avg Freq. This is because these
parameters overlap with those of the power measurement parameters, and the power measurement
parameters are used in their place.
• If any of the power measurement modes is set to ON, cycle mode of automated measurement of waveform
parameters is set to OFF.
Probe Settings (Probe Setup)
The pairs of voltage and current input channels are fixed. Refer to the table below and apply the appropriate
voltage and current signals.
Power Measurement
Voltage Input Channel
Current Input Channel
Power Measurement1
CH1
CH2
Power Measurement2
CH3
CH4
Probe Attenuation Ratio and Voltage-to-Current Conversion Ratio (Probe CH1/CH2 or Probe CH3/CH4)
Select one of the probe attenuation ratios (CH1 or CH3) and one of the voltage-to-current conversion ratios (CH2
or CH4) listed below.
• CH1, CH3
0.001:1, 0.002:1, 0.005:1, 0.01:1, 0.02:1, 0.05:1, 0.1:1, 0.2:1, 0.5:1, 1:1, 2:1, 5:1, 10:1, 20:1, 50:1, 100:1,
200:1, 500:1, 1000:1, 2000:1
• CH2, CH4
0.001 A:1 V, 0.002 A:1 V, 0.005 A:1 V, 0.01 A:1 V, 0.02 A:1 V, 0.05 A:1 V, 0.1 A:1 V, 0.2 A:1 V, 0.5 A:1 V, 1 A:1 V,
2 A:1 V, 5 A:1 V, 10 A:1 V, 20 A:1 V, 50 A:1 V, 100 A:1 V, 200 A:1 V, 500 A:1 V, 1000 A:1 V, 2000 A:1 V
Changing the probe attenuation ratio or voltage-to-current conversion ratio here will change the same ratio in
the CH key probe settings.
17 Power Supply Analysis Feature (Power Analysis and Power Measurement, Option)