4-58
IM DLM3054-01EN
Every Slow CH
The instrument triggers when it detects a slow channel message. To trigger using Every Slow CH, select the
slow channel message type from one of the settings below.
This is selectable when the format version (explained later) is APR2016 or JAN2010.
• Short: Short message
• Enhanced: Enhanced message
Slow CH ID/Data
The instrument triggers on the AND of the slow channel ID and Data conditions.
• Slow CH Type (Slow CH Type)
This item is the same as that of “Every Slow CH” explained earlier.
• Trigger Condition (Condition Setup)
Set the slow channel id/data trigger conditions.
When the message type is Short
• Input Format (Input format)
Set the ID and data pattern input format to Hex (hexadecimal) or Dec (decimal).
• Comparison Condition (Condition)
The data trigger condition is met when the result of comparing the ID and data reference values to the
input signal’s ID and data values meets the specified comparison condition.
• Select whether to compare the ID or data. If you select both, the instrument triggers on the AND
condition.
• Set the comparison conditions for ID and data.
Data
True
1
When the value matches the specified data pattern
False
1
When the value does not match the specified data pattern
ID=a
2
Data=a
2
When the value is equal to the reference value
ID≠a
2
Data≠a
2
When the value is not equal to the reference value
a≤ID
2
a≤Data
2
When the value is greater than or equal to the reference value
ID≤b
2
Data≤b
2
When the value is less than or equal to the reference value
a≤ID≤b
3
a≤Data≤b
3
When the value is within the reference range (including the reference values)
ID<a, b<ID
3
Data<a, b<Data
3
When the value is outside the reference range (excluding the reference values)
1 Set the data pattern for the data nibble in hexadecimal (Hex) or binary (Bin) notation. If the slow channel
type is Short, set an 8 bit data pattern. If slow channel type is Enhanced and the ID and Data message
format (explained later) (
) is “12bit data, 8bit ID,” set a 12 bit data pattern. If the slow
channel type Enhanced and the ID and Data message format is “16bit data, 4bit ID,” set a 16 bit data
pattern.
2 Set one reference value
3 Set two reference values
• Reference Values (a and b)
Set the reference values in Hex (hexadecimal) or Dec (decimal) according to the input format. The
selectable ranges are as follows:
ID
Data
Hex
Dec
Hex
Dec
0 to F
0 to 15
00 to FF
0 to 255
If the comparison condition is a ≤ ID ≤ b, a ≤ Data ≤ b, “ID < a or b < ID,” or “Data < a or b < Data,” the two
reference values are automatically adjusted so that the lower limit is less than or equal to the upper limit.
4 Triggering