App-13
IM DLM3054-01EN
Appendix 6 Trigger and Detected Points of SENT Signals
The following table shows the trigger and detected points of SENT signals.
Trigger/search type
Trigger/detected point
Normal,
with SYNC/CAL error,
with S&C error
Nibble
Number
with error
NDV in S&C
with error
NDV in DataNDV
with error
NDV in CRC
with error
Every Fast CH
End of S&C of the relevant frame
Fast CH S&C
End of S&C of the relevant frame
Not triggered or
searched
End of S&C of the relevant frame
Fast CH Data
End of the last piece of data of the relevant frame
Not triggered when
there is an NDV error
in the data being
compared
End of the last
piece of data
of the relevant
frame
Every Slow CH
End of S&C of the last fast channel of the relevant slow channel
Slow CH ID/Data
Error
Successive CAL
Pulses
(Preferred
Option)
Trigger point: End of S&C of the succeeding frame
Detected point: End of S&C of the relevant frame
Successive CAL
Pulses
(Option 2)
End of S&C of the relevant frame
Nibble Number
Trigger point: End of S&C of the succeeding frame
Detected point: End of S&C of the relevant frame
Nibble Data
Value (NDV)
End of the relevant nibble
Fast CH CRC
End of the relevant CRC
Status and
Communication
(S&C)
End of the relevant S&C
Not triggered or
searched
End of the relevant S&C
Slow CH CRC
End of S&C of the last fast channel of the relevant slow channel
For the figure of the SENT signal frame format, see the explanation of SENT trigger.
Appendix