15-12
IM DLM3054-01EN
ID
Input Format (Input format)
Select the ID input format from one of the settings below.
• Pattern: Set the ID bit pattern in hexadecimal or binary notation.
• Message: Select this setting to set the ID or Data search conditions by using messages and signals that
are defined in the symbol definition file (sbl).
Bit Pattern (Hex/Bin)
Set the 11-bit standard format ID bit pattern or the 29-bit extended format ID bit pattern in hexadecimal (Hex)
or binary (Bin) notation. The ID search condition is met when the input signal ID bit pattern matches the
specified bit pattern.
Remote Frame/Data Frame
Set the search source frame to Remote Frame or Data Frame.
Data
You can set a search condition based on the data field value. Specify the size, position, condition, and data
pattern settings. You can only set these items when you set the trigger source frame to Data Frame.
Comparison Size (Size)
Set the data length to be compared. The data pattern with the specified data length is compared to
the input signal data pattern. If “0” is specified, the instrument assumes that a point has been detected
regardless of the input signal’s data field value.
Selectable range: 0 to 8 bytes
Comparison Start Position (Position)
Set the position to start comparing the data pattern. The left end of the data field is position 0.
Selectable range: 0 to 63 bytes
• If the comparison size (Size) is set to 0, comparison start position (Position) cannot be set.
• The selectable range for position is limited so that Size + Position is less than or equal to 64.
Comparison Condition (Condition)
The data search condition is met when the result of comparing the data pattern or reference value to the
input signal’s data field meets the specified comparison condition.
True
When the value matches the data pattern
False
When the value does not match the data pattern
Data=a
1
When the value is equal to the reference value
Data≠a
1
When the value is not equal to the reference value
a≤Data
1
When the value is greater than or equal to the reference value
Data≤b
1
When the value is less than or equal to the reference value
a≤Data≤b
2
When the value is within the reference range (including the reference values)
Data<a, b<Data
2
When the value is outside the reference range (excluding the reference values)
1 Set one reference value
2 Set two reference values
Data Pattern (Hex/Bin)
If the comparison condition is set to True or False, set the data pattern for the specified size in hexadecimal
(Hex) or binary (Bin) notation.
15 Analyzing and Searching Serial Bus Signals