18-3
IM DLM3054-01EN
Searching the List
You can move to the following record number.
• Delta Max: Record number whose time difference between data triggers is highest
• Delta Min: Record number whose time difference between data triggers is lowest
• Oldest: The oldest record number
• Latest: The most recent record number
Searching History Waveforms (Search)
Search Condition (Condition)
Select the search logic. If you select AND, history waveforms that meet the logical product of search conditions
1 to 4 (described later) will be searched for. If you select OR, those that meet the logical sum of will be searched
for.
• Simple: Searches for history waveforms that enter a rectangular zone
• AND: Searches for history waveforms that meet all search conditions
• OR: Searches for history waveforms that meet any of the search conditions
Search Conditions (1 to 4)
For search conditions 1 to 4*, set the source waveform, search range (the zone or the waveform parameter
upper and lower limits), and the search criterion.
* 1 and 2 on the 2-channel model.
Search conditions 1 to 4 are shared with the reference conditions of GO/NO-GO determination.
Search Criterion (Condition)
Sets whether the source waveform must be in or out of the search range to be detected.
• IN: The source waveform is detected when it is in the search range.
• OUT: The source waveform is detected when it is out of the search range.
Source Waveform (Trace)
You can set the source waveform to one of the waveforms below.
CH1 to CH4/LOGIC,
1, 4
Math1 to Math4, XY1,
2
XY2,
2
FFT1,
3
FFT2
3
1 CH4 or LOGIC, whichever the corresponding key is illuminated, can be selected. If you select LOGIC, select
the
(Bit0 to Bit7).
2 Waveform zones cannot be specified for XY waveforms.
3 Only waveform parameters can be specified for LOGIC and FFT.
Search Range Mode (Mode)
You can set the method of setting the search range to one of the settings below.
• RectZone: Rectangular zone.
• WaveZone: Waveform zone.
• PolygonZone: Polygonal zone.
• Parameter: Sets upper and lower limits for one measured waveform parameter
The methods for setting zones and the upper and lower limits of waveform parameters are the same as with GO/
NO-GO determination (action).
18 Displaying and Searching History Waveforms