13-2
• Resistance temperature detector (RTD) input
Input bias current:
Approx. +1mA (flows from terminal A.)
Burnout indication:
RTD burnout or A-wire burnout . . . . U AL01
B-wire burnout or C-wire burnout. . . U AL01, AL03
2 or more wires burnout . . . . . . . . . . U AL01, AL03
Effect of wiring
resistance:
Max. ±0.05%FS/
Ω
Allowable wiring
resistance:
10
Ω
or less for range No. 53 to 62 (Zener barrier cannot be used.)
85
Ω
or less for ranges other than above range (including the resistance of
the Zener barrier)
• DC voltage input
Input impedance:
Min. 1M
Ω
Input bias current:
1V range or less . . . . . . . . . . . . . .Max. 1µA (flows to the (+) terminal)
0 to 5V, 1 to 5V range . . . . . . . . .Max. 3.5µA (flows to the (+) terminal)
0 to 10V range . . . . . . . . . . . . . . .Max. 7µA (flows to the (+) terminal)
Burnout indication:
Dow AL02
However, the burnout cannot be detected in a range of 0 to 10V.
• DC current input
Input impedance:
Max. 100
Ω
Burnout indication:
Dow AL02
However, the burnout cannot be detected in a range of 0 to 20mA.
Allowable input
current:
Max. 30mA
Allowable input
voltage:
Max.4V(a higher voltage might cause device failure)
●
Motor feedback potentiometer input (R1 model)
Allowable resistance: 100 to 2500
Ω
Burnout detection:
AL07 indication
●
RSP input
Input type:
Linear 0 to 20mA/4 to 20mA or linear 0 to 5V/1 to 5V/0 to 10V
Scaling:
Possible in a range of -1999 to +9999. It is also possible to set the decimal
point position.
Sampling cycle:
100ms
Indication accuracy:
±0.1%FS±1digit (at an ambient temperature of 23±2˚C)
• Voltage input specifications
Input impedance:
Max. 1M
Ω
Input bias current:
0 to 5V, 1 to 5V range . . . . . . . . . . Max. 3µA (flows to the (+) terminal)
0 to 10V range . . . . . . . . . . . . . . . . Max. 5µA (flows to the (+) terminal)
Burnout indication:
Dow AL06
• Current input specifications
Input impedance:
Max. 100
Ω
Burnout indication:
Dow AL06
However, the burnout cannot be detected in a range of 0 to 20mA.
Chapter 13. SPECIFICATIONS
Содержание SDC35
Страница 16: ......
Страница 17: ......
Страница 19: ... display key para key or V key key enter key 2 second press hold of para key V ...
Страница 32: ......
Страница 46: ......
Страница 58: ......
Страница 71: ......
Страница 78: ...5 20 Chapter 5 DETAILED DESCRIPTION OF EACH FUNCTION Item Bank Display Contents Initial value User level ...
Страница 84: ......
Страница 95: ......
Страница 172: ......
Страница 174: ...6 2 Chapter 6 LIST OF DISPLAYS AND SETTING DATA ...
Страница 211: ......
Страница 230: ......
Страница 240: ......
Страница 248: ......
Страница 260: ......
Страница 268: ......
Страница 270: ......
Страница 279: ......