X - R i t e e X a c t ™ I n s t r u m e n t
29
Measurement Condition
Used to select the measurement condition for colorimetric (e.g. CIE L*a*b*) and densitometric
(e.g. Density, TVI, Trapping, etc.) functions. Options are:
•
M0(No) – No Filters
•
M1(D50)
•
M2(UVC) – UV Cut Filter
•
M3(Pol) – Polarization Filter
Illuminant/Observer
Used to select the illuminant and observer combination.
∆
E Method
Used to select the delta E method for color measurements. Options are: CIE
∆
E* (1976),
∆
ECMC,
CIE
∆
E* (2000), and CIE
∆
E* (1994). When
Δ
ECMC, CIE
∆
E* (2000), or CIE
∆
E* (1994), you can
adjust the belonging constants.
Density Status
Used to select the status for all densitometric functions. Options are: Status A, E, I, T, and G.
Density White Base
Used to set a paper or absolute requirement for various density functions. lf you select Paper, as
white base, all subsequent measurements will refer exclusively to the printed ink layer, while the
paper color effect will not be included. lf you select Absolute, besides the printed ink layer, the
measurement will also include the paper effect.
To select a paper measurement requirement for all densitometric functions, select the Paper
check box. To select no paper requirements, select the Absolute check box. Individual
requirements can also be set for each densitometric function by selecting the Custom button. The
functions include Density, All Densities, Density Trend, Density & L*a*b*, CMY Balance, Contrast,
Hue Error & Grayness, Ink QC, and Gray Balance QC.
Density Format
Used to select the density value display format. Options are: Normal (x.xx) and High (x.xxx).
Ink Order (Auto Patch tools only)
Used to set the process ink order for the Auto Patch Tool.
Tap the up and down arrow buttons to set the order.