8 Hygienic reprocessing
8.2 Intervals
8.3 Preparing hygienic reprocessing
Requirement
•
The device is switched off
(see "4.6 Switching the device off",
.
•
The device is disconnected from the patient.
1. Disconnect the device from the power supply.
2. Remove accessories from the device.
3. Dismantle the reusable breathing circuit into its individual parts
(see "8.4 Disassembly of the reusable breathing circuit",
page 176)
4. If necessary: Dismantle the accessories into individual parts.
5. Dispose of all disposables correctly
Result
All parts have been prepared for hygienic reprocessing.
Part
Interval
After each use
At least 1 x weekly
After transportation
of an infected
patient or upon
expiry of the filter's
service life (at least
every 6 months)
All parts (except the
hygiene filter)
x
x
-
Hygiene filter
-
-
x