4 Preparation and operation
4.7.13 Performing ventilation in RSI mode
Requirement
•
The device is switched on.
•
MEDUtrigger is connected to the device.
•
RSI
was set as the ventilation mode for a new patient
(see "4.7.3 Selecting a ventilation mode for a new patient",
page 69)
. The Demand function is automatically activated.
1. Select the
RSI mode
field using the right-hand navigation
knob.