7 Description of the modes
7.3.3 CPAP mode
The set value CPAP/PEEP is used to increase the pressure level of
respiration in order to raise the functional residual capacity (FRC)
of a spontaneously breathing patient. The patient is able to
breathe spontaneously without any restriction at the set pressure
level. The CPAP mode is used exclusively on patients with adequate
spontaneous breathing.
In principle, the pressure is set at the end of expiration (PEEP).
The set maximum ventilation pressure (pMax) ensures the safety of
the patient.
Description
Abbreviation
CPAP
Long form
Continuous Positive Airway Pressure
Type
Pressure-controlled
Requirement
None
Ventilation parameters
Left-hand navigation knob
-
Central navigation knob
-
Right-hand navigation knob
•
PEEP
•
pMax
•
Emergency mode
CPAP
Pressure
Time
1/Freq. (spontaneous)
1/Freq. (spontaneous)
pMax
CPAP / PEEP
CPAP