3 Description
3.5.3 Patient valve (reusable breathing circuit)
3-7 Patient valve (reusable breathing circuit)
Measuring circuit
The device measures the patient's vital parameters
via the measuring circuit. The measuring circuit
comprises:
•
Measuring circuit connector
•
PEEP control tube
•
Pressure measuring tube
•
CO
2
measuring tube (only with Capnography
option)
Water filter (only with Capnography
option)
The water filter protects the measuring chamber of
the device against moisture and contamination from
the patient's respiratory gas.
Measuring circuit connector
Connects the measuring circuit to the connection for
the measuring circuit on the device.
No.
Designation
Description
1
2
3
4
5
No.
Designation
Description
Control cover
Together with the PEEP control diaphragm,
this creates a pressure chamber for PEEP control.
PEEP control diaphragm
Together with the control cover, this creates a pressure
chamber for PEEP control.
Main body
Provides a connection for a mask, endotracheal tube
or the elbow.