7 Description of the modes
If this inspiratory pressure is achieved (pInsp = pMax - 5 mbar),
the device administers as much volume as possible. If this
volume deviates from the set tidal volume, the device triggers
the medium-priority alarm
Vt not achievable
.
7.3.8 SIMV mode
Description
Abbreviation
SIMV
Long form
Synchronized Intermittent Mandatory
Ventilation
Type
Volume-controlled
Requirement
SIMV option is activated
Ventilation parameters
Left-hand navigation knob
Vt
Central navigation knob
Freq.
Right-hand navigation knob
•
PEEP
•
pMax
•
I:E
•
Emergency mode
Pressure
Time
SIMV
1/Freq. (set)
Mechanical
ventilation
Synchronized
mechanical
ventilation
1/Freq. (current)
Trigger time slot
'
T
Spontaneous respiration
Patient
trigger
PEEP
p
Max