2 Safety
2.3.4 Ventilation
Warning
Risk of injury due to lack of patient monitoring!
If the patient is not supervised during ventilation, delayed
responses of medical personnel to alarms and error messages may
result in serious injuries to the patient.
Always monitor patients during ventilation.
Be sure to react immediately to alarms and error messages as
well as a deterioration in the condition of the patient.
Risk of injury from condensate in the FlowCheck sensor and
the patient valve at temperatures below 5 °C!
With longer term ventilation of patients at temperatures below
5 °C, the moisture from expiratory breath can condense in the
FlowCheck sensor and patient valve. This may interfere with the
functioning of the parts and injure the patient.
Quickly transfer the patient to a warmer location.
At temperatures below 5 °C use a breathing system filter to
extend the period of application.
Risk of poisoning due to an overly high concentration of
oxygen during ventilation!
Highly concentrated oxygen can have a toxic effect on the patient
if administered for too long and depending on the age of the
patient.
Do not use highly concentrated oxygen on a patient for too
long during ventilation.
Do not use the device for the ventilation of premature babies
(born before the end of the 36th week of pregnancy).
Risk of injury due to ventilation pressures which are too high
or too low!
Ventilation pressures which are too high or too low may result in
injury to the patient.
Check correct ventilation on the display.
Adjust the maximum ventilation pressure (pMax) to suit the
connected patient.
Risk of injury due to alarm limits which are too high or too
low!
Alarm limits which are either too high or too low can prevent the
device from emitting an alarm, thereby putting the patient at risk.
Always set alarm limits which have been adapted to the
patient.