4 Preparation and operation
Result
The device is operated in oxygen or Air Mix mode.
4.7.7 Performing inhalation (only with Inhalation
option)
Requirement
•
The patient is not connected via an endotracheal tube.
•
An inhalation adapter is connected
.
•
The device is switched on.
•
The start menu is on the display.
1. Select the
New patient
field.
2. Select the height and gender
(see "14.2 Calculation of body
weight on the basis of body height", page 258)
Oxygen mode is activated as standard for all emergency modes.
Using a nebulizer prevents treatment in Inhalation mode!
The device is not suitable for nebulizers in inhalation mode.
The device does not create sufficient pressure for this function.
Do not use nebulizers in inhalation mode with this device.