116
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
5 User menu
In this menu you can set the scale for the time axis of the etCO
2
trend. The following setting options exist: auto, 5 min, 10 min,
30 min, 60 min, 120 min. At a time setting of 5 minutes or
10 minutes, the device records a determined value every
15 seconds. At a setting of 30, 60 or 120 minutes, an average
value is recorded every 30 seconds.
In the "auto" setting, the x axis is scaled automatically depending
on the duration of the application.
The unit of the etCO
2
trend is based on the CO
2
unit. This can
be selected in the operator menu under the menu item Device
configuration
(see "6.3.7 Device configuration", page 129)
5.3.7 Bluetooth
®
(only with Bluetooth
®
data
transmission option)
5-8 Submenu Bluetooth
®
If the
Bluetooth
®
data transmission
option is unlocked and
activated in the operator menu
(see "4.15 Unlocking options",
, you can activate and deactivate the Bluetooth
®
connection here. If the Bluetooth
®
connection is activated, an
operation documentation system can connect to the device to
retrieve session data.
Содержание MEDUMAT Standard2
Страница 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Страница 110: ...110 EN MEDUMAT Standard2 WM 68071 05 2021 5 User menu 5 3 Settings in the user menu 5 3 1 Mode 5 2 Submenu Mode...
Страница 269: ......
Страница 270: ......
Страница 271: ......