4 Preparation and operation
3. Place MEDUtrigger between the mask and the following end of
the breathing circuit closest to the patient:
•
Patient valve
or
•
FlowCheck sensor
or
•
Connector with CO
2
connection
or
•
Elbow
If you use a breathing system filter, always place MEDUtrigger
between the mask and the breathing system filter.
If you are not using MEDUtrigger (e.g., with CPAP applications),
release it from the breathing circuit and fix to the breathing circuit
with the hook and loop strap with clip.