8 Hygienic reprocessing
5. After the specified exposure time, rinse the parts for at least
10 s and textiles for at least 2 min with drinking-water quality
in order to remove all of the disinfectant residue.
6. Allow the parts to dry fully.
7. Check the parts for residue and any remaining dirt.
8. If dirt is visible: Repeat the cleaning and disinfection.
Result
The parts are immersion disinfected.
8.7.1 Immerse disinfecting the reusable measuring
circuit
The principle described only applies to the following parts of the
reusable measuring circuit:
•
Pressure measuring tube
•
PEEP control tube
•
Measuring circuit connector
Requirement
•
The reusable measuring circuit is disconnected from the
reusable breathing circuit
reusable breathing circuit", page 176)
•
The reusable measuring circuit has been cleaned manually.
1. The disinfectants, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page 196)
2. Prepare the immersion disinfection solution in accordance with
the specifications of the disinfectant agent manufacturer.
3. Connect a sterile disposable syringe (20 ml) to one end of a
tube.
Risk of infection and contamination due to insufficient
hygienic reprocessing of the measuring circuit!
Rinsing the measuring circuit in alternating directions does not
guarantee any bacteria reduction and may injure the patient.
Only rinse measuring tubes in one direction.