126
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
6 Operator menu
Export configuration
The
Export configuration
function allows you to export all the
configuration settings performed on the device to an SD card.
Import configuration
The
Import configuration
function allows you to import the
configuration settings exported to an SD card from a device onto
a second device.
Following the import, the second device is configured in exactly the
same way as the original device. The access code for the operator
menu is also adopted.
If you do not wish to adopt the customer-specific password for the
operator menu, you have two options:
•
Before the export: Reset the password to
000000
and export
the configuration.
•
Prior to import: Set the password to
000000
before exiting the
operator menu.
When exporting, all the configuration settings (including the
options) are transferred with the exception of the following
configuration settings:
•
Date and time
•
Serial number
•
Device runtime
•
Filter runtime
•
Date of last function check
•
Date of last maintenance
•
Number of start-ups
Configuration imports are saved in the service data.
Configurations can only be transferred between devices with
the same software version. Options subject to a charge are
only imported if these options are already activated.