4 Preparation and operation
4.16 Updating the software
Requirement
•
The device is connected to the line power.
•
A fully charged battery is inserted in the device.
•
The operator menu has been called up
(see "6.1 Navigating the operator menu", page 120)
1. If necessary: Download the software from the Login area of
the WEINMANN Emergency website to the SD card.
2. If the software is available as a ZIP file: Unzip the software.
The software is available in the folder as a file named
WM28981-x.x.hex
.
3. Place the file in the SD card's root directory.
When doing so, note: The file for the software update must not
be in a sub-folder.
4. Select the menu item
Software update
.