Using the ventilator
35
EN
Table 1:
Audio response up to
serial number 7999
Meaning
“Open oxygen cylinder”
Open oxygen cylinder valve
slowly
.
“Check respiration and
select mode”
Depending on whether or not the patient is breathing, set
MEDUMAT Easy to one of the modes Demandflow (Page 27), mask
ventilation or intubation (Page 24).
“Adjust settings”
Depending on patient weight, set respiration frequency and minute
volume (Page 23).
“Connect patient”
Connect patient to ventilator via ventilation hose and patient valve
using the patient mask or the connector of the tracheal tube.
“Demandflow mode”
Demandflow mode is selected.
“Mask ventilation mode”
“Tilt head back”
“Secure mask tightly”
Mask ventilation mode is selected.
While tilting the head back, use the E-C technique to seal the mask.
“Tube ventilation mode”
“Ventilation pressure limit
45 mbar”
Tube ventilation mode is selected.
Maximum ventilation pressure for tube ventilation.
“Check airways and minute
volume”
MEDUMAT Easy has measured excessive airway resistance. Check the
airways or adjust respiratory frequency and minute volume settings to
suit the patient (Page 23).
“Device malfunction”
“Administer alternative
ventilation”
The device is faulty or the battery is failing. The device can no longer be
used for ventilation. You must therefore use another ventilation method
(Page 39).
“Check pressure hose
system and gas supply”
MEDUMAT Easy has measured low pressure on the inlet side. Check
whether the O
2
cylinder still contains sufficient oxygen and that the
oxygen hose is not leaking, kinked or jammed.
“Rule out respiratory arrest
and check mask fit”
MEDUMAT Easy can no longer detect a breathing pulse (trigger) in
Demandflow mode. Check the patient’s breathing, and if necessary
switch to a different ventilation mode. Check the connections and
mask fit.
“Close oxygen cylinder”
After switching off the ventilator, turn off the O
2
cylinder or the external
O
2
supply.
“Check ventilation system
and settings“
Disconnection: a pressure rise of 8 mbar is not achieved during the
inspiration phase under controlled ventilation. This is usually due to an
interruption of the ventilation system or to a low minute volume setting.
Check the connections or adjust the minute volume to suit the patient.
“Selected language:
English” (Deutsch, Français,
...)
When selecting the language for the audio response, press the mask/
tube switch as many times as necessary until the desired language is
announced.
“Audio response is off”
Confirmation that audio response is deactivated.