64
Technical data
EN
Subject to technical change without
notice.
MEDUMAT Easy
Alarm sound
pressure
60 dB (A)
Pressure gauge
accuracy
±5% of upper range
value
Patient valve
resistance, reusable:
Inspiration
Expiration
Spont. breathing
<6 mbar
(3)
at 60 l/min
<6 mbar
(3)
at 60 l/min
1.5 mbar
(3)
at 30 l/min
Resistance of patient
hose system,
disposable (as per
EN 794-3):
Inspiration
Expiration
Spont. breathing
22,4 mbar
(3)
at 60 l/min
3,46 mbar
(3)
at 60 l/min
1,54 mbar
(3)
at 30 l/min
Elasticity of breathing
system
negligible
Patient valve
dead space
12,8 ml (reusable)
8 ml (disposable)
Materials used for
disposable hose sys-
tem
PC, silicone, PVC, PP, PS,
EVA, K-Resin
®
Components with
critical flow direction
Patient valve
Components
containing latex
None
Demandflow mode:
– trigger
– peak flow
– shut-off pressure
< 1 mbar
(3)
> 40 l/min
3 mbar
MEDUMAT Easy
Recommended safe distances between portable or mobile HF
telecommunication devices (e.g. mobile phones) and MEDUMAT Easy
Rated output of HF
device
in W
Safe distance depending on transmission frequency
in m
80 MHz - 800 MHz
800 MHz – 2.5 GHz
0.01
0.12
0.23
0.1
0.34
0.73
1
1.20
2.30
10
3.40
7.30
100
12.00
23.00