Functional checks
45
EN
8. Functional checks
Before each use, after each dismantling and reassem-
bly, and at the very least every 6 months, the user must
carry out functional checks on the ventilator.
Note:
Before carrying out the functional check on
MEDUMAT Easy, you must connect the ventilation
hose and the patient valve.
MEDUMAT Easy must not be used if the functional
checks reveal defects or deviations from the speci-
fied parameters.
First try to correct the error with the help of the in-
formation provided in section “9. Troubleshooting”
on page 54. If this is not possible, have the unit re-
paired by the manufacturer – WEINMANN
Emergency – or by specialists explicitly authorized by
WEINMANN Emergency.
A complete functional check includes:
•
“8.2 Checking for leaks in the system” on
page 47;
•
“8.3 Checking patient hose system” on
page 48;
•
“8.4 Checking the minute volume” on page 49;
•
“ Checking the breath volume” on page 49;
•
“8.5 Checking maximum ventilation pressure”
on page 50;
•
“8.6 Checking Demandflow” on page 51;
•
“8.7 Checking the alarm systems” on page 52;
We recommend that you always hold reserve stocks
of the following items:
•
replacement washers for the connections;
Содержание Emergency MEDUMAT Easy
Страница 2: ...English 3 Espa ol 70 Portugu s 146...
Страница 5: ...Overview 5 EN B 10 14 13 11 12 15 C 16 17 19 20 18 21 or...
Страница 69: ......