42
Hygienic preparation
EN
7.3 Ventilation hose
Caution!
Only reusable hose system WM 22520 (scope of
supply) is suitable for the hygienic preparation de-
scribed here. Do
not
subject disposable hose system
WM 28110 available as an accessory to hygiene
preparation. Replace it with a new one.
1. Take the ventilation hose and the pressure
gauge hose off both connection ports.
Caution!
Always grasp the hoses
at the end
, as
shown in the drawing, otherwise the hoses may
be damaged or torn off. Close both ends of the
pressure gauge hose.
2. Perform the hygienic preparation as described in
section 7.6.
3. For reassembling, see “5.2 Ventilation hose” on
7.4 Masks
Perform the hygienic preparation of the masks as
described in section 7.6.
7.5 Fittings
Ensure that no liquids get into the connections for
the oxygen fittings, otherwise there is a risk of explo-
sion, especially if you are using products for disinfec-
ting by wiping which contain alcohol.
For external cleaning of fittings (e.g. pressure reduc-
er, valve), use only a clean cloth. The cloth may be dry
or moistened with clean water.