58
Servicing
EN
unit. Use only batteries supplied by
WEINMANN Emergency.
10.2 Change valve membrane in patient valve
If one of the valve membranes in the expiration or
spontaneous breathing arms of the patient valve is
crinkled, sticky or misshapen, it must be changed.
Spontaneous breathing arm
1. Take the spontaneous breathing insert out of
the patient valve. To do so, push the two lock-
ing lugs out of their seat, using a small screw-
driver, for example.
2. Pull the defective valve membrane out of the
spontaneous breathing insert using pointed
tweezers.
3. Put in a new valve membrane.
4. Push the spontaneous breathing insert back into
the patient valve.
Expiration arm
1. Use pointed tweezers to pull the defective valve
membrane out of the expiration arm.
2. Insert a new valve membrane.
10.3 Storage
If you do not intend to use MEDUMAT Easy for a
long period, we recommend the following storage
precautions:
Spontaneous
breathing
arm
Expiration-
arm