Safety instructions
15
EN
•
MEDUMAT Easy should be used only by medi-
cally qualified personnel who have had training
in ventilation techniques. Incorrect use can cause
severe physical injury.
•
Please note that a safe distance must be main-
tained between MEDUMAT Easy and equipment
that emits HF radiation (e.g. mobile phones),
otherwise malfunctions may occur (see “ Rec-
ommended safe distances between portable or
mobile HF telecommunication devices (e.g. mo-
bile phones) and MEDUMAT Easy” on page 64).
•
We recommend that maintenance work such as
inspections and repairs be performed only by the
manufacturer, WEINMANN Emergency, or by
qualified technicians expressly authorized by
WEINMANN Emergency.
•
Malfunctions and a lack of biocompatibility may
result if third-party articles are used. Please note
that in such cases all warranty entitlement and li-
ability claims shall be void where items other
than the accessories recommended in the in-
structions for use or original spare parts are
used.
•
Design changes to the device are not permitted
and may put patients and users at risk.
Oxygen
Spontaneous explosive reactions can occur if high-
pressure oxygen comes into contact with flammable
substances (fat, oil, alcohol etc.):
•
Keep the equipment and all screw connections
absolutely free from oil and grease.
•
Always wash your hands before starting to work
on the oxygen supply.