Using the ventilator
31
EN
Disconnection alarm in Demandflow
mode
If the patient does not trigger MEDUMAT Easy with-
in 15 seconds, the “Disconnection” alarm is given. If
audio response is enabled, the unit announces “Rule
out respiratory arrest and check mask fit”.
< 2.7 bar O
2
alarm
Oxygen pressure at the pressure connection to the
MEDUMAT Easy has dropped to less than 2.7 bar. The
reason is usually an almost empty oxygen cylinder.
In this case MEDUMAT Easy can no longer function
correctly because the operating parameters are no
longer within the permissible limits.
If audio response is enabled, the ventilator announces:
“Check pressure hose system and gas supply”.
alarm
The battery is failing. Failure of the automatic venti-
lation function must be expected. Immediate steps
must therefore be taken to provide alternative venti-
lation (see “6.12 Alternative ventilation proce-
dures” on page 39).
If audio response is enabled, you will hear the mes-
sage: “Device malfunction! Administer alternative
ventilation”.
The ventilator must be switched off before the bat-
tery can be changed (see “ Changing the main bat-
tery” on page 57).
Cancelling acoustic alarm
If there is an alarm, you can temporarily cancel the
acoustic alarm by pressing the alarm mute button:
+ –