20
Installation
EN
Always grasp the ventilation hose and
pressure gauge tube by their end only,
otherwise they may be damaged or split.
3. Connect the patient valve to the other end of the
ventilation hose and pressure gauge tube.
4. If a mask is being used for ventilation, attach the
mask connection to the patient valve (identical
with tube connection),
or
if the patient is intubated, attach the patient
valve to the tube.
Filter
If a filter is to be used, this should be installed
between the patient side connector on the patient
valve and the mask or tube.
Always follow the instructions supplied by the
manufacturer of the filter.
Position
> PSU< 134
°