24
Using the ventilator
EN
The figures shown in the table are only recom-
mendations. Different settings may be required
in cases of pulmonary damage or for special indi-
cations.
To see the relationship between the values, see
diagram “12.2 Relationship between ventilation
parameters” on page 66.
Maximum ventilation pressure
1. Use the mask/tube switch to set the maximum
ventilation pressure. The LEDs light up in active
mode.
If the maximum level is reached, e.g. in cases where
compliance is inadequate, MEDUMAT Easy sets off a
stenosis alarm (see “ Stenosis alarm” on page 30).
Note:
The mask/tube switch can only be operated when
the ventilator is switched on.
Recommended maximum ventilation pressure:
Device
Mask ventilation
Intubation
20 mbar
45 mbar
Audio response up
to serial number
7999
Audio response
enabled: “Mask
ventilation mode. Tilt
back head, secure
mask tightly”
Audio response
enabled: “Tube
ventilation mode.
Ventilation pressure
limit 45 mbar”
From serial number
8000 and following
repair/servicing/
firmware update
Audio response
enabled: “Ventilation
pressure limit
20 mbar”
Audio response
enabled: “Ventilation
pressure limit
45 mbar”