Technical data
EN
Demandflow
An inspiration impulse (trigger) at V2 opens valves
V3 and V4.
Electronic system
The microprocessor-controlled electronic system sets
the ventilation parameters and monitors ventilation,
and also O
2
supply and power supply. If necessary, a
visual and acoustic alarm is given. The ventilator has
an audio response facility that can be switched on
for user guidance.
12.2 Relationship between ventilation parameters
The following diagram shows the relationship be-
tween the ventilation parameters “minute volume”
and “respiratory frequency”:
Minute volume (MV) in l/min
Frequenc
y
in
l/mi
n
Minute volume (V
t
) in l
1,1
0,1
0,3
0,6
0,9
1,1
1,1