Functional checks
51
EN
MEDUMAT Easy must set off the
stenosis
alarm.
Usually this occurs after the second inspiration
stroke.
5. Switch the mask/tube switch to
.
6. Repeat the check for intubation ventilation with
the settings:
–
MV: 9 l/min / frequency: 10 min
-1
–
P
max
:
(45 mbar)
If audio response is enabled, the ventilator must
announce “Tube ventilation mode, ventilation
pressure limit 45 mbar” (from serial number
8000 and following repair/servicing/ firmware
update: “Ventilation pressure limit 45 mbar”).
During this test you must not assist the expira-
tion stroke. This ensures that the pressure builds
up slowly. Between 40 and 50 mbar
MEDUMAT Easy must set off the
stenosis
alarm.
Usually this occurs after the second inspiration
stroke.
7. Switch off MEDUMAT Easy again.
8.6 Checking Demandflow
1. Make sure that MEDUMAT Easy is switched off
and the oxygen cylinder is open.
2. Attach the test bag to the patient valve with the
adapter from test set WM 15323.
3. Select the setting “Demandflow”.
4. Switch on MEDUMAT Easy. The green “De-
mandflow” LED lights up.
If audio response is enabled, the ventilator an-
nounces “Demandflow mode”.
5. Simulate an inspiration pulse by hand by firmly
squeezing and quickly releasing the test bag.