Using the ventilator
23
EN
During this test, the four LEDs in alarm panel
flash on and off and a short alarm tone sounds.
If an error is found, all the LEDs in alarm panel
continue flashing and an alarm sounds. If this hap-
pens, MEDUMAT Easy must not be used for venti-
lation.
If audio response is enabled, you will hear the
message “Device malfunction! Administer alter-
native ventilation”.
After the self test, the ventilator repeatedly
checks the oxygen cylinder pressure until ade-
quate pressure is detected. Otherwise an alarm is
sounded.
The MEDUMAT Easy will then start to function
with the selected ventilation settings.
6.2 Selecting the ventilation settings
We recommend selection before switching on, to
prevent unnecessary waste of oxygen.
Respiratory frequency and minute
volume
1. Set the respiratory frequency and the minute
volume with regulator knob.
Stenosis
Disconnection
< 2,7 bar O
2
Recommended ventilation settings:
yellow orange
brown
Body weight
10 kg
30 kg
60 kg
80 kg
110 kg
Respiratory
frequency
30 min
-1
16 min
-1
11 min
-1
10 min
-1
10 min
-1
Minute
volume
3 l/min
5 l/min
7 l/min
9 l/min
11 l/min
13
14
0,3
0,6
1,1
16
15
13
14
14
12
12
11
10
10
9
11
16
7
5
30
3
MV (l/min)
Freq.(min
-1
)
Demand-
flow