36
Proper operating procedures
To ensure your system runs efficiently, follow the operating procedures and
guidelines in
“Starting the detector” on page 27
Spare parts
Replace only parts mentioned in this document. For spare parts details, see
the Waters Quality Parts Locator on the Waters Web site’s Services & Support
page.
Recommendations:
•
For optimal baseline stability, keep the detector door closed at all times.
•
Filter solvents to prolong column life, reduce pressure fluctuations, and
decrease baseline noise.
•
To conserve lamp life, extinguish the lamp while leaving the detector
running but idle. Note, however, that you should do so only when the
lamp will remain extinguished more than 4 hours.
•
If you use a mobile phase with additives, flush it from the detector before
powering-off to prevent these adverse conditions:
–
Plugging of solvent lines and the flow cell
–
Damaging of instrument components
–
Microbial growth
Caution:
• To ensure optimum performance of the flow cell, ensure that eluent is
flowing before powering-on the detector. If, however, you must
power-on the detector before the eluent is flowing, extinguish the
lamp after establishing communications.
• To avoid damaging the detector or column, remove the column and
disconnect the detector before you flush the system.