Error messages and troubleshooting
67
Error messages preventing operation
During initialization and operation, the detector can display “<Error>” in the
control panel, signifying a usually terminating malfunction and preventing
further operation of the detector.
When you encounter such an error, ensure that
•
the flow cell is clean;
•
the front door is shut securely.
Cycle power to the detector. If the terminating error persists, contact Waters
Technical Service.
Vcc fuse failed
Fuse failure was
detected.
1. Cycle power to the
detector.
2. Contact Waters
Technical Service.
Instrument error messages:
Error Message
Description
Corrective Action
Both lamp and flow cell
memory devices are
blank.
Neither the lamp nor
the flow cell memory
device has been
programmed.
1. Cycle power to the
detector.
2. If problem persists,
contact Waters
Technical Service.
Communications with
the Dsp failed. Dsp
firmware is not running.
Personality card
electronic problem.
1. Cycle power to the
detector.
2. If problem persists,
contact Waters
Technical Service.
Startup and operating error messages: (Continued)
Error Message
Description
Corrective Action