
4
Calculating absorbance
The detector computes absorbance by subtracting the dark current (see
) and reference spectrum from the acquired spectrum.
Absorbance is based on the principles of Beer’s law.
Beer’s law
The relationship between the quantity of light of a particular wavelength
arriving at the photodiode and the concentration of the sample passing
through the flow cell is described by the Beer-Lambert law (commonly called
Beer’s law).
Beer’s law is expressed as:
A
=
lc
where
A
= dimensionless quantity measured in absorbance units
= constant of proportionality, known as the molar extinction coefficient
l
= path length, in centimeters (1.0 cm in the detector’s normal flow cell)
c
= concentration, in moles per liter
Beer’s law applies only to well-equilibrated dilute solutions. It assumes that
the refractive index of the sample remains constant, that the light is
monochromatic, and that no stray light reaches the detector element. As
concentration increases, the chemical and instrumental requirements of
Beer’s law can be violated, resulting in a deviation from (absorbance versus
concentration) linearity.
Содержание ACQUITY UPC2
Страница 84: ...Index 4...