Maintaining the detector
49
14. Connect the identification cable on the front of the detector to the
identification connection on the front of the flow cell.
15. Close the front panel door.
16. Before you power-on the detector, prime the system.
Result:
Doing so fills the flow cell with liquid carbon dioxide, removing
air from the cell.
Replacing the lamp
Waters warrants 2000 hours of lamp life for the lamp, or one year since date of
purchase, whichever comes first The software senses the ACQUITY UPC
2
PDA lamp when you install it, and records its serial number and installation
date in the lamp change record table.
Change the lamp when it repeatedly fails to ignite or when the detector fails
to undergo calibration.
To remove the lamp:
1. Power-off the lamp:
•
To power-off the lamp manually, click PDA Detector in the left-hand
pane of the console, and then click the lamp icon.
Tip
: The green LED on the console darkens as does the Lamp LED
on the door.
•
To power-off the lamp using a timed event, see the instructions in
the Empower or MassLynx online Help.
Warning:
To prevent burn injuries, allow the lamp to cool for 30 minutes
before removing it. The lamp housing gets extremely hot during
operation.
Warning:
To avoid eye injury from ultraviolet radiation exposure
• power-off the detector before changing the lamp;
• wear eye protection that filters ultraviolet light;
• keep the lamp in the housing during operation.