52
4. Gently push the lamp forward until it bottoms into position.
5. Ensure the lamp is flush against the optics bench.
6. Tighten the two captive screws, and then reconnect the lamp power
connector.
7. Power-on the detector, and then wait about 1 hour for the lamp to warm
before resuming operations.
Tip:
Cycling power to the detector (that is, powering-off and then
powering-on the instrument) initiates the verification procedures.
Result:
The lamp’s serial number and installation date are
automatically entered into the Lamp Change Record table.
Reading lamp energy
Recommendation:
Impurities in the flow cell can affect reading lamp energy.
Ensure that the flow cell is clean before you read lamp energy.
To read the lamp energy:
1. In the console, select PDA Detector from the system tree.
2. In the PDA detector information window, click Maintain > Read energy
> Read. The Read Energy dialog box appears.
3. Click Close.
Replacing the fuses
The detector requires two 100 to 240 Vac, 50 to 60 Hz, F 3.15-A, 250-V FAST
BLO, 5 × 20 mm (IEC) fuses.
Caution:
To prevent the lamp from binding and ensure that it is
properly seated in the lamp housing, alternate between tightening
the captive screws and pushing the lamp forward.
Warning:
To avoid electric shock, power-off and unplug the
detector before examining the fuses. For continued protection
against fire, replace fuses only with the required type and rating.