70
Detector troubleshooting
Detector troubleshooting:
Symptom
Possible cause
Corrective action
Both LEDs unlit
No power
1. Inspect line cord
connections.
2. Test electrical outlet
for power.
Open (spent) or
defective fuse
Replace fuse (see
).
Change in reference
spectrum
Air bubbles trapped in
flow cell
• Reseat and check
alignment of flow
• Flush the flow cell
(see
), or
apply slight
backpressure on the
detector waste
outlet.
• Ensure the
backpressure
regulator is
connected to the
detector waste
outlet.
Communication
problems
Configuration problem
Check Ethernet
configuration.
Improper or defective
Ethernet cable
Replace the cable with
a shielded Ethernet
cable.