48
9. Gently push the front of the assembly until it seats on the front
alignment pin.
Installing the flow cell assembly:
10. Continue to insert the flow cell until the three thumbscrews align with
their holes in the optics bench.
11. Using the 1/4-inch, flat-blade screwdriver, finger-tighten the
thumbscrews, and then secure them.
12. Connect the tubing from the column outlet to the flow cell inlet.
Finger-tighten, then use the right-angle tool (nut extender) to tighten
1/4 turn.
13. Connect the outlet tubing from the outlet of the flow cell to the vent
valve located in the ACQUITY UPC
2
Convergence Manager.
Finger-tighten, then use the right-angle tool (nut extender) to tighten
1/4 turn.
Caution:
To prevent the flow cell from binding and ensure that it is
properly seated in the optics bench, alternate between tightening the
captive screws and pushing the flow cell forward.
TP03457
Dowel alignment pin
Содержание ACQUITY UPC2
Страница 84: ...Index 4...