Error messages and troubleshooting
69
Shutter failed to home
Shutter could not home. 1. Flush the flow cell.
2. Establish flow.
3. Reset
communications from
the control panel or
console.
4. Contact Waters
Technical Service
Temperature controller
A/D failed
Personality card
electronic error.
Contact Waters
Technical Service to
replace the personality
card.
Temperature sensor
hardware fault
Sensor error.
1. Replace the
temperature sensor.
2. If the problem
persists, contact
Waters Technical
Service to replace the
personality card.
Thermal controller
disabled
Thermal controller was
disabled.
1. Cycle power to the
detector.
2. Contact Waters
Technical Service.
Wavelength outside of
allowed range
Instrument method
usage error.
1. Modify the
instrument method.
Ensure wavelength
end is lower than 800
nm.
Wavelength verification
failure
Instrument failed
power-up initialization
wavelength verification
step.
1. Establish flow and
reset the instrument.
2. If step 1 fails,
perform erbium
calculation.
Instrument error messages: (Continued)
Error Message
Description
Corrective Action