50
2. Power-off the detector, and disconnect the power cable from the rear
panel.
Alternative:
To save time, leave the detector powered on for 15 minutes
after you power-off the lamp. Doing so will allow the fan to blow cool air
on the lamp, cooling it faster. Be sure to power-off the detector, and
disconnect the power cable from the rear panel after 15 minutes has
elapsed.
3. Allow the lamp to cool for 30 minutes (or 15 minutes with the fan
running), and then open the door, gently pulling its right edge toward
you.
4. Detach the lamp power connector from the detector.
Warning:
To avoid burn injuries, wait 30 minutes (or 15 minutes
with the fan running) before touching the lamp or its housing.
Warning:
To avoid electric shock, power-off and unplug the
detector before detaching the lamp power connector from the
detector.
Caution:
To avoid damaging the detector’s electronics, power-off
and unplug the detector before detaching the lamp power
connector from the detector.
Содержание ACQUITY UPC2