Maintaining the detector
53
Suspect a faulty fuse when
•
the detector fails to power-on;
•
the cooling fan does not operate.
Requirement:
Replace both fuses, even when only one is faulty.
To replace the fuses:
Requirement:
Replace both fuses, even when only one is open or otherwise
defective.
1. Power-off the detector, and disconnect the power cord from the power
entry module.
2. Pinch the sides of the spring-loaded fuse holder, which is above the
power entry module on the rear panel of the detector. With minimum
pressure, withdraw the spring-loaded fuse holder.
3. Remove and discard the fuses.
4. Make sure that the new fuses are properly rated for your requirements,
and then insert them into the holder and the holder into the power entry
module, gently pushing until the assembly locks into position.
5. Reconnect the power cord to the power entry module.
Cleaning the instrument’s exterior
Use a soft cloth, dampened with water, to clean the outside of the detector.
TP02523
Fuses
Fuse holder
Power entry
module