Maintaining the detector
47
b. Enter zero flow in the composition field or press the Stop Flow icon
and enter a flow rate of zero. The system will vent to atmospheric
pressure through the vent valve in the ACQUITY UPC
2
Convergence Manager.
3. Open the detector front panel door.
4. Disconnect the detector’s inlet and outlet tubing and the flow cell
identification cable from their appropriate connections.
ACQUITY UPC
2
PDA detector flow cell
5. Using a 1/4-inch flat-blade screwdriver, loosen the three captive
thumbscrews on the flow cell assembly’s front plate.
6. Grip the flow cell, and gently pull it toward you.
7. Unpack and inspect the new flow cell.
8. Square the flow cell assembly in front of the opening, and then insert it
into the optics bench.
Note:
Use the alignment pin on the optics bench as a guide to installing
the flow cell.
TP03458
Flow cell ID connector
Flow cell handle
Lamp
Flow cell
Tubing
outlet
Tubing
inlet
Thumbscrews