Installing a manual injector
3-7
The following figure shows the inject-start connection from the I/O terminal
block on the pump to connector A on a 2489 detector.
Inject start connection to a 2489 detector
Requirements:
•
Set the manual injector in the LOAD position when you start the data
control software. The busLAC/E card can fail to boot if the injector is left
in the INJECT position.
•
When you use the manual injector in conjunction with compositional
gradients, turn the injector back to the LOAD position before the end of
the run. Leaving the injector in the INJECT position automatically
triggers the gradient table for subsequent injections, but data
acquisition does not start.
For instructions on operating the FlexInject Dual Manual Injector with a
detector, refer to the
Waters FlexInject Dual Manual Injector Module
Installation Guide
.
1500-series pump
I/O terminal block
2489 detector
I/O terminal block
(connector A)
Inject
−
2
1
1 Inject Start +
2 Inject Start
−