Powering off the pump
4-23
Powering off the pump
To power-off the pump:
1. If you have been using a buffer, purge it from the pump and other HPLC
system components with HPLC-grade water.
2. If the pump will be idle for more than one day, purge with a
methanol/water solution to prevent the growth of microorganisms.
3. If the pump is running, use a shutdown method to ramp the flow to zero.
4. Turn the power switch to the O (OFF) position.
Caution:
Since abrupt flow changes can damage columns, it is
advisable to ramp flow to 0 using an appropriate shutdown
method. Refer to your data control software’s online Help for
details.