5-12
Maintaining the HPLC Pump
Cleaning the plunger
To clean the plunger:
1. Remove the plunger from the plunger assembly.
2. Clean the plunger by sonicating it in 50:50 methanol/water for a few
minutes.
Assembled plunger
Inspecting the plunger
After cleaning the plunger, inspect it for damage by holding it under a bright
white light and looking down its length for nicks and scratches. It is easier to
see scratches under a bright light than to feel them with your fingers.
•
If the plunger is not scratched or otherwise damaged, reassemble it with
new seals. Continue with
“Replacing the Plunger” on page 5-12
.
•
If the plunger is damaged, replace both the plunger and the seals.
Continue with
“Replacing the Plunger” on page 5-12
Replacing the Plunger
To replace the plunger:
1. Reassemble the components of the plunger assembly as shown in the
figure above.
2. Lubricate the ball and seat with high-pressure grease.
Make sure the brass seat is located at the bottom of the piston.
Snap-ring
Plunger
Washer
Spring
Retainer
Plunger
Retainer seal
Piston
Retainer
Spring
1515/1525 plunger assembly
1525µ plunger assembly