Maintaining the HPLC Pump
Required materials
•
Open-end wrench, 5/16-inch (startup kit)
•
Allen wrench, 5/32-inch (startup kit)
•
Allen wrench, 9/64-inch (startup kit)
•
Adjustable wrench (startup kit)
•
Snap-ring pliers (for 1515/1525 pumps only) (part number WAT025263)
•
Plunger insertion tool and seal extraction tool (startup kit)
•
Sonicator
•
HPLC-grade water
•
HPLC-grade methanol
•
Replacement plunger
•
Replacement plunger seals
Removing the plunger
To remove the plunger from the pump:
1. Remove the pump head as described on
2. Remove the four head-support screws with the 9/64-inch Allen wrench,
and then carefully slide the head support assembly and the indicator rod
off the pump.
Exposed head support assembly
3. Set the head support assembly on the benchtop.
4. Use the data control software to set the flow rate to 0.3 mL/min and
initiate pump flow.
Head support
screws
Head support
screws
1515/1525 head support assembly
1525µ head support assembly