SERVICE MANUAL
Troubleshooting
(Rev. 1)
Rotograph EVO
8-67
•
The image shows TMJ at different height
This fact is described in Figure 8-7.
Figure 8-7
Possible cause:
The mid sagittal plane is not vertical.
This can be a patient’s
problem, but if the defect is always present, check the laser beam.
Remedy:
Verify that the laser beam is vertical; this check can be performed
very quickly using the laser beam and verifying that falls on the
center of the chin support; remove the chin support itself and check
that the beam falls on the center of the holes used to fix the chin
itself.
If not, a possible cause can be due to a non perfect horizontality of
the chin arm, that must be adjusted using the relevant screws.
Содержание Rotograph EVO
Страница 1: ...Release 14 September 2011 Rev 1 Rotograph EVO 0051 Service Manual ...
Страница 3: ...SERVICE MANUAL Revision history Rotograph EVO Rev 1 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 19: ...SERVICE MANUAL Safety information Rotograph EVO Rev 1 2 10 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 32: ...SERVICE MANUAL Description Rev 1 Rotograph EVO 3 13 Figure 3 1 Block diagram ...
Страница 37: ...SERVICE MANUAL Description Rotograph EVO Rev 1 3 18 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 51: ...SERVICE MANUAL Technical data Rotograph EVO Rev 1 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 59: ...SERVICE MANUAL Pre installation Rotograph EVO Rev 1 5 8 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 169: ...SERVICE MANUAL Troubleshooting Rotograph EVO Rev 1 8 78 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 171: ...SERVICE MANUAL Schematics and drawings Rotograph EVO Rev 1 9 2 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 219: ...SERVICE MANUAL Spare Parts Rotograph EVO Rev 1 10 10 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 222: ......