SERVICE MANUAL
Troubleshooting
(Rev. 1)
Rotograph EVO
8-65
•
Anterior teeth are enlarged and blurred
The Figure 8-5 depicts the resulting film.
Figure 8-5
Possible cause:
In the case the patient is positioned too backward.
Remedy:
Check the patient positioning.
If, after the correct patient’s positioning, the problem still remain,
check the alignment of rotating arm.
A more accurate check can be performed using the service tool and
verifying the distance between the two lateral metal parts.