SERVICE MANUAL
Troubleshooting
Rotograph EVO
(Rev. 1)
8-38
8.3.4.2. Y
Axis
zero
This function allows the set of the offset of Y axis, the motion along mid
sagittal plane of the patient, in order to place the correct position of the
central path. Once the Patient centering (>0<) key is pressed, the
following message is displayed:
Y
Z
E
R
O
[ Y
Y
] a
Z E R O O F F S
±±±±
x x x x x
where:
•
"YY" is the stepper motor speed, that can be changed using the
parameter increase/decrease keys.
•
"a" is the status of the zero optical barrier; 1 means engaged (i.e.
barrier interrupted), 0 otherwise.
•
"xxxxx" is the current value of the offset, that can be changed with
the left/right keys.
Содержание Rotograph EVO
Страница 1: ...Release 14 September 2011 Rev 1 Rotograph EVO 0051 Service Manual ...
Страница 3: ...SERVICE MANUAL Revision history Rotograph EVO Rev 1 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 19: ...SERVICE MANUAL Safety information Rotograph EVO Rev 1 2 10 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 32: ...SERVICE MANUAL Description Rev 1 Rotograph EVO 3 13 Figure 3 1 Block diagram ...
Страница 37: ...SERVICE MANUAL Description Rotograph EVO Rev 1 3 18 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 51: ...SERVICE MANUAL Technical data Rotograph EVO Rev 1 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 59: ...SERVICE MANUAL Pre installation Rotograph EVO Rev 1 5 8 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 169: ...SERVICE MANUAL Troubleshooting Rotograph EVO Rev 1 8 78 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 171: ...SERVICE MANUAL Schematics and drawings Rotograph EVO Rev 1 9 2 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 219: ...SERVICE MANUAL Spare Parts Rotograph EVO Rev 1 10 10 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 222: ......